Configuring the DS2
DS2 Docking Station
186
INDUSTRIAL SCIENTIFIC
Version 9.1 (P/N: 17112798)
Step Instruction
10.
Click the OK button from the Edit Docking Station dialog box to save your changes.
11.
The gas cylinder is now ready to be used.
CAUTION:
When configuring Gas In connections, be careful that you have configured the
correct gas type in the DSSAC that matches the gas type in the cylinder connected to the Gas In
connection on the IDS. If this setup is incorrect, your calibrations may be run with the incorrect
gas, rendering the results inaccurate.
Table 6-4. Gas Inlets Tab Icons
Icon Description
Explanation
Solid Green Circle
The cylinder is ok or is using fresh air.
Solid Yellow Circle
The cylinder is low.
Solid Red Circle
The cylinder is empty.
Yellow Circle with a Rex X
The cylinder will expire within 30 days.
Red Circle with a White X
The cylinder is expired.
Yellow Circle with an
exclamation point
The Gas In connection is disabled.
NOTE:
If you hover your mouse pointer over one of the icons on the Gas Inlets tab, the status
of the cylinder or Gas In connection displays.
6.7. Adding Gas from Industrial Scientific
To add an ISC gas, follow the instructions below.
Step Instruction
1.
From the Edit Docking Station dialog box, click on the Gas Inlets tab.
2.
Select the tab corresponding to the Gas In connection that you are configuring.
3.
Click the Assign button and select ISC.
4.
The Cylinder Configuration dialog box appears.