DS2 Docking Station
Event Scheduling
Version 9.1 (P/N: 17112798)
INDUSTRIAL SCIENTIFIC
221
Run Time:
12:00 AM
Calibration
Interval:
Monthly
Day:
1
st
Effective Date:
1/1/03
Run Time:
12:00 AM
Download Datalog
Interval:
Upon Docking
Effective Date:
1/1/03
Alarms Interval:
Upon
Docking
Run Time:
After Download Datalog
Effective Date:
1/1/03
9.3. Special
Events
Special Events are custom events that you can assign to specific instruments. If an instrument is
assigned to a Special Event, the Global Event no longer applies to that instrument. You may
wish to setup special events to handle groups of instruments that have special maintenance
requirements. For example, you may have a group of heavily used instruments that you wish to
calibrate weekly instead of monthly, as specified in the global event.
You can add the following types of Special Events:
Bump Test
Bump Test (Tango)
Diagnostics
Calibration
Alarms
Download datalog.
NOTE:
The “Bump Test (Tango)” special event is system created. It applies to any
Tango instrument that is in single-sensor mode or has two installed sensors and one is in
calibration fail. When any Tango TX1 instrument is in either state, DSS will override any
other bump test event for the unit.
To add a Special Event for an instrument, follow the instructions listed below.
Step Instruction
1.
Open the DSSAC application.
2.
Expand the Events option in the navigation pane, if necessary.
3.
Expand the Instrument option that is underneath Events, if necessary.
4.
Click on Special. Any instrument Special Events that have been previously added
appear in the contents pane. The contents pane displays the Event ID, the Type of
event, the current scheduled Interval, and the Event Name. If the Event ID column
contains an icon with a red line through it, the event is disabled.