Configuring Instruments
DS2 Docking Station
188
INDUSTRIAL SCIENTIFIC
Version 9.1 (P/N: 17112798)
Element Description
Response Factor Name
- This is a customizable name for the
custom response factor. The maximum length allowed entered is 16
characters.
Response Value
- This is the response value for the custom
response factor. Each item in the list has a checkbox. Clicking in
the checkbox toggles the check on and off. This can be a number
from 0 to 99.99 inclusive.
Double-clicking on an item in the list toggles the check on and off.
“Checking” a custom response factor adds it to the Favorite
Response Factor list. “Un-checking” a custom response factor
removes it to the Favorite Response Factor list.
Edit Custom Button
The Edit Custom button is used to edit the selected custom response
factor. Clicking the edit Custom button opens the Edit Custom
Response Factor form. The Edit Custom Response Factor form is
used to modify the name and or value of the custom response value.
Remove Custom Button The Remove Custom button is used to delete the selected custom
response factor. When a Custom response factor is removed, then
any installed PID sensor currently configured to use that custom
response factor is automatically reverted back to Isobutylene.
Favorite Response
Factors list
The Favorite Response Factor list displays the selected response
factors, standard or custom, which have been deemed a favorite. A
maximum of 5 responses, in any combination of standard or
custom, may be selected. The order in which the favorite responses
are selected is the order in which they are saved to the instrument
Clear Favorites Button
The Clear Favorites button is used for removing the checks from the
selected responses, thus removing them from the Favorite