![Industrial Scientific DS2 Docking Station Скачать руководство пользователя страница 234](http://html1.mh-extra.com/html/industrial-scientific/ds2-docking-station/ds2-docking-station_administrators-manual_2054246234.webp)
Event Scheduling
DS2 Docking Station
224
INDUSTRIAL SCIENTIFIC
Version 9.1 (P/N: 17112798)
NOTE:
The next time that you view the special event, the
Journal
tab will be visible. The
Journal tab contains records of the last time that the event ran for various instruments. See
Chapter 10 for more information about the Journal.
9.4. DS2 Defaults for Scheduled Events (v5.0 and higher)
For new DSS installs, the table below shall be the default schedules for all events. For upgrades
to previous versions of DSS, the default schedule is not altered by the installer. For upgrades, all
current event schedules remain unmodified by the installer. See the table below.
Table 9-2. DS2 Defaults for Scheduled Events
Event Priority
Schedule
Time
New in
v5.0
Enabled
IDS Settings Read
1
Daily
12:00am
No
Yes
IDS Settings Update
2
Daily
12:00am
No
Yes
IDS Diagnostics
3
Daily
12:30am Yes
Yes
Instrument Settings Read
4
Upon Docking
1:00am Yes
Yes
Instrument Settings Update
5
Upon Docking
1:00am Yes
Yes
Instrument Diagnostics
6
Upon Docking
1:30am Yes
Yes
Bump Test
7
Daily
2:00am Yes Yes
Calibration 8
Monthly
2:00am Yes Yes
Datalog Download
9
Upon Docking
2:30am Yes
Yes
Alarm Events Download
10
Upon Docking
3:00am Yes
Yes