DS2 Docking Station
Configuring
Instruments
Version 9.1 (P/N: 17112798)
INDUSTRIAL SCIENTIFIC
193
Figure 5-17. Sample Note for Exchanging Instruments
The following settings will be copied from the old instrument to the new one:
Access code
Backlight setting
Language
Location
Recording interval
Status
TWA timebase
All options
All users
All sites.
Copying of sensor settings occurs according to the following rules.
When the replacement instrument has a sensor of the same sensor type in the same
position as the original, the sensor settings will be copied.
If the replacement instrument is missing 1 or more sensors that were present in the
original instrument, the settings will still be copied for the sensors that are present.
If the replacement instrument has a sensor of a type which was not present in the original
instrument, the settings of that sensor will not be changed.
If the replacement instrument has 1 sensor which is the same type as the original, but it is
in a different position, the settings for that sensor will still be copied.
If the replacement instrument has 2 sensors of same type, 1 in the same position as the
original instrument, and 1 in a different position, only the sensor in the correct position
will be modified.