DS2 Docking Station
Configuring
Instruments
Version 9.1 (P/N: 17112798)
INDUSTRIAL SCIENTIFIC
205
Figure 5-25. The Bump Test Tab of the Edit Instrument Dialog Box with Sample Results
Step Instruction
7.
You can sort the bump test results by clicking on the column heading by which you
would like to sort.
8.
The following information is displayed for each bump test.
Item Description
Serial Number
The serial number of the sensor
Time
The date and time when the bump test was performed
Gas Type
The type of gas that was used for the bump test
Passed/Failed
Whether or not the sensor passed the bump test. (Failed
entries also display in red.)
Bump Threshold
The bump threshold for the bump record. If no value has been
recorded, the column will display the default value of 50.
Bump Timeout
The bump timeout for the bump record. If no value has been
recorded (pre-5.0 docking stations), the column will display
“N/A”, since some older docking stations bump with a timeout
of two minutes, and others at 90 seconds.
9.
To view the details of a bump test, double-click on its entry in the list.
Alternative: You can also right-click on the entry and select Properties from the
context menu.