9.5 Communication Methods
111
9
Standard Event Status Register (SESR)
The Standard Event Status Register is an 8-bit register.
If any bit in the Standard Event Status Register is set to 1 (after masking by the Standard
Event Status Enable Register), bit 5 (ESB) of the Status Byte Register is set to 1.
See:
"Standard Event Status Register (SESR) and Standard Event Status Enable Register (SESER)" (p.
112)
The Standard Event Status Register is cleared in the following situations:
• When a
CLS
command is executed
• When an event register query (
ESR?
) is executed
• When the instrument is powered on
Event Registers
Bit 7
PON
Power-On Flag
Set to 1 when the power is turned on, or upon recovery from an outage.
Bit 6
URQ
User Request
unused
Bit 5
CME
Command error (The command to the message terminator is ignored.)
This bit is set to 1 when a received command contains a syntactic or semantic
error:
•
Program header error
•
Incorrect number of data parameters
•
Invalid parameter format
•
Received a command not supported by the instrument
Bit 4
EXE
Execution Error
This bit is set to 1 when a received command cannot be executed for some
reason.
•
The specified data value is outside of the set range
•
The specified setting data cannot be set
•
Execution is prevented by some other operation being performed
Bit 3
(unused)
DDE
Not used by this instrument
Device-Dependent Error
This bit is set to 1 when a command cannot be executed due to some reason
other than a command error, a query error or an execution error.
Bit 2
QYE
Query Error (the output queue is cleared)
This bit is set to 1 when a query error is detected by the output queue control.
•
When an attempt has been made to read an empty output queue (GP-IB
only)
•
When the data overflows the output queue
•
When data in the output queue has been lost
•
When the next command is received while there is data in the output
queue
Bit 1
RQC
(unused)
Request Control
Bit 0
OPC
Operation Complete
This bit is set to 1 in response to an
OPC
command.
•
It indicates the completion of operations of all messages up to the
OPC
command
Содержание RM3542
Страница 2: ......
Страница 6: ...Contents iv ...
Страница 16: ...Operating Precautions 10 ...
Страница 26: ...1 3 Screen Organization 20 ...
Страница 32: ...2 3 Turning the Power On and Off 26 ...
Страница 46: ...3 8 Confirming Faulty Measurements 40 ...
Страница 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Страница 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Страница 90: ...7 3 Printing 84 ...
Страница 172: ...9 10 Device Compliance Statement 166 ...
Страница 190: ...11 4 Disposing of the Instrument 184 ...
Страница 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Страница 214: ...Index Index 4 ...
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......