9.7 Message Reference
126
Write and Read the Service Request Enable Register (SRER)
Syntax
Command
SRE
<0 to 255
(NR1)>
Query
Response
SRE?
<0 to 255
(NR1)>
Description
Command
The SRER mask is set to the numerical value 0 to 255.
Although NRf numerical values are accepted, values to the right of the decimal are
rounded to the nearest integer.
Bit 6 and unused bits 2, 3 and 7 are ignored.
The data is initialized to zero at power-on.
Query
The contents of the SRER, as set by the
SRE
command, are returned as an NR1
value (0 to 255). Bit 6 and unused bits 2, 3 and 7 always return as zero.
Example
SRE 33
Set SRER bits 0 and 5 to 1.
SRE?
33
SRER bits 0 and 5 have been set to 1.
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
unused
0
ESB
MAV
unused
unused
ESE1
ESE0
Read the Status Byte and MSS Bit
Syntax
Query
Response
STB?
<0 to 255 (NR1)>
Description
The contents of the STB are returned as an NR1 value (0 to 255).
The response message has no header.
Example
STB?
16
STB bit 4 has been set to 1.
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
unused
MSS
ESB
MAV
unused
unused
ESE1
ESE0
Request a Sample
Syntax
Command
TRG
Description
Performs one measurement when external triggering is enabled.
When Statistical Calculation is ON, imports calculation data.
When the memory function is enabled, the measured value is stored.
Example
:TRIGger:SOURce EXTernal;
TRG
Note
When an error occurs with the Settings Monitor function enabled, triggering is disabled (p. 53).
Содержание RM3542
Страница 2: ......
Страница 6: ...Contents iv ...
Страница 16: ...Operating Precautions 10 ...
Страница 26: ...1 3 Screen Organization 20 ...
Страница 32: ...2 3 Turning the Power On and Off 26 ...
Страница 46: ...3 8 Confirming Faulty Measurements 40 ...
Страница 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Страница 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Страница 90: ...7 3 Printing 84 ...
Страница 172: ...9 10 Device Compliance Statement 166 ...
Страница 190: ...11 4 Disposing of the Instrument 184 ...
Страница 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Страница 214: ...Index Index 4 ...
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......