4.10 Retrying Measurement After a Fault (Retry Function)
56
The Retry function causes measurement to be
automatically retried when a measurement fault
occurs due to probe chatter.
During Retry, all measurement operations including Con-
tact Improvement and DELAY2 (but excluding DELAY1)
are restarted.
If a measurement fault persists after the specified contin-
uous retry interval (e.g., if the DUT is not connected),
retrying is aborted and the
EOM
signal is output. When
Retry is enabled, the maximum time to end-of-measure-
ment occurs when recovering from a measurement fault
immediately before the retry interval expires, which
approaches the sum of the retry interval setting plus nor-
mal measurement time. Decreasing test throughput may
indicate probe maintenance is required.
4.10 Retrying Measurement After a Fault
(Retry Function)
Retry
Chatter
Probe Contact
Condition
Start
TRIG
Contact
Improver
Measurement
Current
Contact Check
End of Measurement
EOM
1
Open the Basic Settings screen.
2
Open the System screen.
3
Select whether to enable or disable Retry.
4
(When enabled (ON selected))
Set the continuous retry interval.
5
Return to the Measurement screen.
The Basic Settings screen appears.
The System screen appears.
[SYSTEM]
1
Selection
2
Selection
1
2
Retry disabled (go to step 5)
Retry enabled (default)
Selection
1
2
Setting range: 1 to 50 ms (default: 50 ms)
3
The confirmation screen appears.
Return to the setting screen.
Save setting and return to
previous screen.
Discard setting and return to
previous screen.
Содержание RM3542
Страница 2: ......
Страница 6: ...Contents iv ...
Страница 16: ...Operating Precautions 10 ...
Страница 26: ...1 3 Screen Organization 20 ...
Страница 32: ...2 3 Turning the Power On and Off 26 ...
Страница 46: ...3 8 Confirming Faulty Measurements 40 ...
Страница 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Страница 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Страница 90: ...7 3 Printing 84 ...
Страница 172: ...9 10 Device Compliance Statement 166 ...
Страница 190: ...11 4 Disposing of the Instrument 184 ...
Страница 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Страница 214: ...Index Index 4 ...
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......