6.2 Store as soon as Measurement is Stable (Auto-Memory Function)
71
6
This function automatically stores the value measured each time the probes contact the sample with
internal triggering. When the specified number of values has been acquired, auto-storage operation
stops.
Statistical calculations are applied to the stored data, with results output to the screen or printer (RS-
232C).
See:
"6.3 Performing Statistical Calculations on Measured Values" (p. 74)
"Chapter 7 Printing" (p. 79)
Data storage and printing can be automatically controlled by the Auto-Memory function.
The Auto-Memory function is disabled by default.
Enable the Auto-Memory function before setting the number of values to store.
Enabling the Auto-Memory function affects other functions as follows:
• Statistical calculation is forced on.
• The voltage level monitor function is forced off (although the setting itself is not set to OFF, the function is
actually disabled).
• The trigger source setting is forced to internal (INT).
When the trigger source is set to external (EXT), the Auto-Memory function is disabled by force.
Deleting Stored Data
Stored data is automatically erased at the following times:
6.2
Store as soon as Measurement is Stable
(Auto-Memory Function)
Printing (p. 82)
Prepare the printer (p. 79).
Enable Auto-Memory and set
the number of values to store.
Set decision criteria
(p. 34).
Measure
Beeper notifies when the specified number of values is stored.
•
when the memory function setting (including auto-mem-
ory) is changed (p. 142)
•
when the range is changed (p. 31)
•
when changing comparator settings (p. 34)
•
when the power is turned off
•
when printing the statistical calculations (p. 82)
•
when the DUT is changed (p. 28)
•
upon system reset (p. 67)
•
upon setting the auto-memory number of values to store
(p. 73)
Содержание RM3542
Страница 2: ......
Страница 6: ...Contents iv ...
Страница 16: ...Operating Precautions 10 ...
Страница 26: ...1 3 Screen Organization 20 ...
Страница 32: ...2 3 Turning the Power On and Off 26 ...
Страница 46: ...3 8 Confirming Faulty Measurements 40 ...
Страница 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Страница 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Страница 90: ...7 3 Printing 84 ...
Страница 172: ...9 10 Device Compliance Statement 166 ...
Страница 190: ...11 4 Disposing of the Instrument 184 ...
Страница 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Страница 214: ...Index Index 4 ...
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......