
3.6 Zero Adjustment
32
When four-terminal measurement (Kelvin connection) is impractical such as when measuring very
small samples, the additional inherent resistance of the two-terminal wiring should be canceled out.
The zero-adjustment function can cancel out up to 10
additional resistance.
Before Zero Adjustment
• The guaranteed accuracy of the instrument applies to four-terminal connections without zero adjustment.
When using four-terminal connections, do not execute zero adjustment. Executing zero adjustment with
incorrect wiring may amplify measurement error. However, zero adjustment may be needed even with four-
terminal measurements if they are affected by a large offset voltage, such as due to thermal emf (LOW
POWER OFF, in 100
to 100 M
ranges).
• Execute zero adjustment when the ambient temperature has changed, or when a probe is replaced.
Execute zero adjustment after the warm-up period following power on.
3.6
Zero Adjustment
1
Open the Basic Settings screen.
2
Select the internal (INT) trigger mode.
3
Short the probes together.
4
Confirm that the measured value does not exceed 10
.
If no measured value is displayed, increment the measurement range (p. 31).
5
Select whether to enable or disable zero adjustment.
After confirming that the measured value does not exceed 10
, execute zero adjustment.
6
Return to the Measurement screen.
The Basic Settings screen appears.
1
Selection
2
Internal trigger
Press the up/down cur-
sor keys to change the
setting.
1
Selection
2
Disable zero adjustment (cancel)
Execute zero adjustment
Содержание RM3542
Страница 2: ......
Страница 6: ...Contents iv ...
Страница 16: ...Operating Precautions 10 ...
Страница 26: ...1 3 Screen Organization 20 ...
Страница 32: ...2 3 Turning the Power On and Off 26 ...
Страница 46: ...3 8 Confirming Faulty Measurements 40 ...
Страница 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Страница 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Страница 90: ...7 3 Printing 84 ...
Страница 172: ...9 10 Device Compliance Statement 166 ...
Страница 190: ...11 4 Disposing of the Instrument 184 ...
Страница 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Страница 214: ...Index Index 4 ...
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......