9.7 Message Reference
143
9
Read Memory Data
Syntax
Query
:MEMory:DATA?
Response
<Measurement value (NR3)>,<Measurement value (NR3)>,...,<Measurement value
(NR3)>
Description
Measured values transferred from memory are separated by commas (“,”). The number of stored
measurement values to be exported can be acquired by the
:MEMory:COUNt?
query.
Data stored by both the data (MEM) and auto-memory (AUTO) modes can be acquired by this
command (although only one mode can be enabled at a time).
See:
"Measurement Value Formats" (p. 140)
Note
•
Memory data is available only by remote command.
•
Memorize measurement values with the TRIG
signal,
F4
[MANU] key on the [MENU]-[TRG]
selection screen, or
TRG
command.
•
Binary data is returned when the Binary sending format is selected.
Set and Query the Memory Point (maximum data quantity)
Syntax
Command
:MEMory:POINt
<1 to 30000>
Query
Response
:MEMory:POINt?
<1 to 30000>
Description
Sets the number of measured values that can be stored.
When the number of stored values reach the memory point number, no further values are stored
(no overwriting occurs).
Example
:MEM:POIN 1000
:MEM:POIN?
1000
Note
•
When the auto-memory mode is enabled, the maximum memory point number is 99. Although
the value may be set higher with the
:MEMory:POINt
command, the memory is limited to 100
measurement values.
•
When the auto-memory mode is enabled, stored data is erased when the memory point number
is changed.
Содержание RM3542
Страница 2: ......
Страница 6: ...Contents iv ...
Страница 16: ...Operating Precautions 10 ...
Страница 26: ...1 3 Screen Organization 20 ...
Страница 32: ...2 3 Turning the Power On and Off 26 ...
Страница 46: ...3 8 Confirming Faulty Measurements 40 ...
Страница 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Страница 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Страница 90: ...7 3 Printing 84 ...
Страница 172: ...9 10 Device Compliance Statement 166 ...
Страница 190: ...11 4 Disposing of the Instrument 184 ...
Страница 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Страница 214: ...Index Index 4 ...
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......