
3.1 Pre-Operation Inspection
27
3
See "Measurement Flow" (p. 2) for an outline of the measurement process from preparation to end-
of-measurement.
Measurement Settings
(Basic Measurements)
Chapter 3
3.1
Pre-Operation Inspection
Do not use the instrument if damage is
found, as electric shock or short-circuit
accidents could result. Contact your
dealer or Hioki representative.
Metal Exposed
Is the power cord insulation torn, or is
any metal exposed?
1
Before using the instrument for the first time, verify that it operates normally to ensure that no dam-
age occurred during storage or shipping. If you find any damage, contact your dealer or Hioki repre-
sentative.
Peripheral Device Inspection
Is the insulation on a measurement
cable torn, or is any metal exposed?
No Metal Exposed
If there is any damage, measured val-
ues may be unstable and measure-
ment errors may occur. Replace the
cable with an undamaged one.
Metal Exposed
No Metal Exposed
If damage is evident, request repairs.
Yes
Is damage to the instrument evident?
Instrument Inspection
When turning power on
Does the self-test screen appear
(model no., version no.)? (p. 25)
No
2
The power cord may be damaged, or
the instrument may be damaged inter-
nally. Request repairs.
Does the Measurement screen appear
after self-test?
No
Yes
The instrument may be damaged inter-
nally. Request repairs.
See:
"11.1 Troubleshooting" (p. 179)
"11.3 Error Displays and Reme-
dies" (p. 181)
An error indication
occurs (ERR)
Yes
Inspection complete
Please read the "Operating Precautions" (p. 7) before use.
Содержание RM3542
Страница 2: ......
Страница 6: ...Contents iv ...
Страница 16: ...Operating Precautions 10 ...
Страница 26: ...1 3 Screen Organization 20 ...
Страница 32: ...2 3 Turning the Power On and Off 26 ...
Страница 46: ...3 8 Confirming Faulty Measurements 40 ...
Страница 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Страница 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Страница 90: ...7 3 Printing 84 ...
Страница 172: ...9 10 Device Compliance Statement 166 ...
Страница 190: ...11 4 Disposing of the Instrument 184 ...
Страница 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Страница 214: ...Index Index 4 ...
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......