9.7 Message Reference
145
9
Set and Query Integration Time (in seconds)
Syntax
Command
:RESistance:APERture
<Range>
,
<Speed>
,
<Integration time
(seconds)
>
Query
Response
:RESistance:APERture?
<Range>
,
<Speed>
<Integration time (seconds)>
Command
:LPResistance:APERture
<Range>
,
<Speed>
,
<Integration time
(seconds)
>
Query
Response
:LPResistance:APERture?
<Range>
,
<Speed>
<Integration time (seconds)>
<Range>
=
(:RESistance) RNG100MIL/ RNG1000MIL/ RNG10/ RNG100/ RNG1000/ RNG10K/
RNG100K/ RNG1000K/ RNG10MEG/ RNG100MEG
(:LPResistance) RNG1000MIL/ RNG10/ RNG100/ RNG1000
<Speed>
= SLOW/ MEDium/ FAST
<Integration time>
= 0.0001 to 0.1 (seconds)
Description
Sets the integration time, in seconds.
Set and Query Contact Improver Operating Mode
Syntax
Command
:RESistance:CIMProve
<Range>
,
<OFF/ HOLD/ PULSe>
Query
Response
:RESistance:CIMProve?
<Range>
<OFF/ HOLD/ PULSE>
Command
:LPResistance:CIMProve
<Range>
,
<OFF/ HOLD/ PULSe>
Query
:LPResistance:CIMProve?
<Range>
Response
<OFF/ HOLD/ PULSE>
<Range>
=
(:RESistance) RNG100MIL/ RNG1000MIL/ RNG10/ RNG100/ RNG1000/ RNG10K/
RNG100K/ RNG1000K/ RNG10MEG/ RNG100MEG
(:LPResistance) RNG1000MIL/ RNG10/ RNG100/ RNG1000
Description
Sets Contact Improver operation to OFF, HOLD or PULSE.
Set and Query Contact Improver Level
Syntax
Command
:RESistance:CIMProve:LEVel
<Range>
,
<Level>
Query
Response
:RESistance:CIMProve:LEVel?
<Range>
<Level>
Command
:LPResistance:CIMProve:LEVel
<Range>
,
<Level>
Query
Response
:LPResistance:CIMProve:LEVel?
<Range>
<Level>
<Range>
=
(:RESistance) RNG100MIL/ RNG1000MIL/ RNG10/ RNG100/ RNG1000/ RNG10K/
RNG100K/ RNG1000K/ RNG10MEG/ RNG100MEG
(:LPResistance) RNG1000MIL/ RNG10/ RNG100/ RNG1000
<Level>
=
L1/ L2/ L3/ L4 (applied current)
(L1: 17 mA/ L2: 25 mA/ L3: 35 mA/ L4: 50 mA)
Description
Sets the level of current applied by the Contact Improver function.
Содержание RM3542
Страница 2: ......
Страница 6: ...Contents iv ...
Страница 16: ...Operating Precautions 10 ...
Страница 26: ...1 3 Screen Organization 20 ...
Страница 32: ...2 3 Turning the Power On and Off 26 ...
Страница 46: ...3 8 Confirming Faulty Measurements 40 ...
Страница 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Страница 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Страница 90: ...7 3 Printing 84 ...
Страница 172: ...9 10 Device Compliance Statement 166 ...
Страница 190: ...11 4 Disposing of the Instrument 184 ...
Страница 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Страница 214: ...Index Index 4 ...
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......