9.7 Message Reference
130
(4) Calculation
Clear and Query the Statistical Calculation State
• A data sample can be taken by the following methods:
1. Press the
F4
[MANU] key on the [MENU]-[TRG] selection screen (p. 70)
2. Apply a TRIG signal to the external I/O connector.
3. Send a
TRG
command.
• The
:CALCulate:STATistics:STATe
command does not clear calculation results.
• When the valid data count is 0,
n-1 returns 0.
• When cleared, the Statistical Calculation function is not turned OFF.
• The upper limit of Cp and CpK is 99.99. If Cp or CpK exceeds 99.99, the value 99.99 is returned.
Statistical Calculation Execution State
Syntax
Command
:CALCulate:STATistics:STATe
<1/ 0/ ON/ OFF>
Query
Response
:CALCulate:STATistics:STATe?
<ON/ OFF>
Example
:CALC:STAT:STAT ON
:CALC:STAT:STAT?
ON
Clear Statistical Calculation Results
Syntax
Command
:CALCulate:STATistics:CLEar
Queries the data count
Syntax
Query
Response
:CALCulate:STATistics:NUMBer?
<Total data count (NR1)>,<Valid data count (NR1)>
Data count = 0 to 30000
Example
:CALC:STAT:NUMB?
23456,23449
Query the Mean value
Syntax
Query
Response
:CALCulate:STATistics:MEAN?
<Mean (NR3)>
Query the Maximum value
Syntax
Query
Response
:CALCulate:STATistics:MAXimum?
<Maximum value (NR3)>,<Data No. of Maximum value (NR1)>
Example
:CALC:STAT:MAX?
12.4859E+3,1124
Query the Minimum value
Syntax
Query
Response
:CALCulate:STATistics:MINimum?
<Minimum value (NR3)>,<Data No. of Minimum value (NR1)>
Содержание RM3542
Страница 2: ......
Страница 6: ...Contents iv ...
Страница 16: ...Operating Precautions 10 ...
Страница 26: ...1 3 Screen Organization 20 ...
Страница 32: ...2 3 Turning the Power On and Off 26 ...
Страница 46: ...3 8 Confirming Faulty Measurements 40 ...
Страница 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Страница 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Страница 90: ...7 3 Printing 84 ...
Страница 172: ...9 10 Device Compliance Statement 166 ...
Страница 190: ...11 4 Disposing of the Instrument 184 ...
Страница 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Страница 214: ...Index Index 4 ...
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......