Chapter 10 Specifications
172
(18) Clock
(19) Power Line Frequency Setting
(20) Reset Functions
(21) Self-Calibration Function
(22) Memory Function
(23) Auto-Memory Function
Auto calendar, auto leap year, 24-hour clock
Accuracy
Approx. 4 minutes/month
Default state
01/01/2009, 00:00:00
Backup battery life
Approx. 10 years (from factory shipping)
Operation
Selects the line voltage frequency
Default state
AUTO (auto-detect upon power on and resetting)
Setting
AUTO (50 or 60 Hz, fixed)/ 50 Hz / 60 Hz
Reset
Operation
Resets settings (except the clock) to factory defaults
RST (Remote command)
Operation
Resets settings (except the clock and interface) to factory defaults
When power is restored, reverts to the settings before
RST was sent
Operation
Compensates for offset voltage and gain of measurement circuit
Compensation timing
When setting is changed and once every 10 minutes
Operation
Measured values are stored by an EXT I/O TRIG signal or by pressing the F4 [MANU]*
button
* Displayed when EXT is selected on the [MENU] – [TRG] selection screen
Default state
OFF
Setting
ON/ OFF
Memory capacity
30,000 (volatile memory, no backup)
Operation
When stable, measured values are acquired by internal continuous triggering during
manual measurement
The beeper sounds when the specified number of measurements is acquired
Memory is cleared when acquired data has been transferred or printed by RS-232C
During Auto-Memory operation, the statistical calculation function is always enabled.
The voltage level monitor function is disabled.
Default state
OFF/ 10
Setting
ON/ OFF
Number of measurements stored: 1 to 99
Содержание RM3542
Страница 2: ......
Страница 6: ...Contents iv ...
Страница 16: ...Operating Precautions 10 ...
Страница 26: ...1 3 Screen Organization 20 ...
Страница 32: ...2 3 Turning the Power On and Off 26 ...
Страница 46: ...3 8 Confirming Faulty Measurements 40 ...
Страница 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Страница 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Страница 90: ...7 3 Printing 84 ...
Страница 172: ...9 10 Device Compliance Statement 166 ...
Страница 190: ...11 4 Disposing of the Instrument 184 ...
Страница 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Страница 214: ...Index Index 4 ...
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......