Chapter 10 Specifications
175
10
RS-232C
Communication contents
Remote control, measured value output (export)
Transfer method
Asynchronous, Full du
plex
Transmission speed
9,600bps (default setting)/ 19,200bps/ 38,400bps
Data length
8 bit
Stop bit
1
Parity
none
Delimiter
Transmit CR+LF, Receive CR or CR+LF
Handshaking
No X-flow, no hardware flow
Protocol
Non-procedure
Connector
Male 9-pin D-sub, with #4-40 attachment screws
Printer
Printing Contents
Measurement data (PRINT signal input), Total data count, Mean, Minimum value (sam-
ple no.), Maximum value (sample no.), Standard deviation of sample, Population stan-
dard deviation, Process capability, No. of IN, No. of Hi and Lo, No. of OvrRng, No. of
Measurement faults
Communications
Communications method
Asynchronous RS-232C
Transmission speed
9,600bps
Data length
8 bits
Stop bit
1 bit
Parity
none
Delimiter
Transmit CR+LF, Receive CR or CR+LF
Handshake
No X-flow, no hardware flow
GP-IB interface (RM3542-01 only)
Communication contents
Remote control
Interface Functions
SH1
All Source Handshake functions are supported.
AH1
All Acceptor Handshake functions are supported.
T6
Basic talker functions are supported.
Serial poll function are supported.
No talk-only mode.
The talker cancel function with MLA (My Listen Address) is supported.
L4
Basic listener functions are supported.
No listen-only mode.
The listener cancel function with MTA (My Talk Address) is supported.
SR1
All Service Request functions are supported.
RL1
All Remote/Local functions are supported.
PP0
No Parallel Poll function.
DC1
All Device Clear functions are supported.
DT1
All Device Trigger functions are supported.
C0
No Controller functions are supported.
Miscellaneous
Conforms to IEEE 488.2
Settings Monitor terminal (SET MONITOR)
Connector
Male 9-pin D-sub, with #4-40 attachment screws
Connection cable
Crossover
Содержание RM3542
Страница 2: ......
Страница 6: ...Contents iv ...
Страница 16: ...Operating Precautions 10 ...
Страница 26: ...1 3 Screen Organization 20 ...
Страница 32: ...2 3 Turning the Power On and Off 26 ...
Страница 46: ...3 8 Confirming Faulty Measurements 40 ...
Страница 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Страница 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Страница 90: ...7 3 Printing 84 ...
Страница 172: ...9 10 Device Compliance Statement 166 ...
Страница 190: ...11 4 Disposing of the Instrument 184 ...
Страница 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Страница 214: ...Index Index 4 ...
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......