6.1 Storing Data at Specific Times (Data Memory Function)
70
Measured values are stored in the instrument’s internal memory according to the following timings.
(up to 30,000 points)
• Every time a measurement is performed by external (EXT) triggering
• When a trigger is applied during internally (INT) triggered measurement
The following three storage methods are available:
• Store upon receiving an EXT I/O TRIG signa
l
• Store upon receiving a
TRG
command (p. 126)
• Store by pressing the
F4
[MANU] key on the [MENU] – [TRG] setting screen.
• This function can only be enabled by remote control. The data memory function should be enabled by
remote control beforehand. This setting is not available from the front panel.
• Stored memory data cannot be viewed on the instrument’s screen. Use remote control commands to export
stored data.
Data Memory Function Operating Procedure
Stored data is automatically erased at the following times:
6.1
Storing Data at Specific Times
(Data Memory Function)
1
Enable data memory mode.
Send this remote command to enable the data memory function:
2
Store measured values.
Execute external trigger measurement, or apply a trigger during internally triggered measurement.
3
Export the stored data.
Send this remote command to export the measured values stored in the instrument:
4
Clear measurement data from instrument memory.
Send this remote command to erase the data from instrument memory:
:MEMory:CLEar
•
when the memory function setting (including auto-mem-
ory) is changed (p. 142)
•
when the range is changed (p. 31)
•
when changing comparator settings (p. 34)
•
when printing the statistical calculations (p. 82)
•
when the DUT is changed (p. 28)
•
upon system reset (p. 67)
Содержание RM3542
Страница 2: ......
Страница 6: ...Contents iv ...
Страница 16: ...Operating Precautions 10 ...
Страница 26: ...1 3 Screen Organization 20 ...
Страница 32: ...2 3 Turning the Power On and Off 26 ...
Страница 46: ...3 8 Confirming Faulty Measurements 40 ...
Страница 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Страница 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Страница 90: ...7 3 Printing 84 ...
Страница 172: ...9 10 Device Compliance Statement 166 ...
Страница 190: ...11 4 Disposing of the Instrument 184 ...
Страница 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Страница 214: ...Index Index 4 ...
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......