6.4 Auto-Exporting Measured Values (at End of Measurement) (Data Output Function)
77
6
At the end of measurement, the measured value is exported to a computer via RS-232C.
See:
"Chapter 9 Communications (RS-232C/ GP-IB Interface)" (p. 97)
6.4
Auto-Exporting Measured Values (at End of
Measurement) (Data Output Function)
• Set the interface to [RS232C] beforehand. This function is not applicable to the GP-IB Interface.
See:
"9.4 Configuring the Communications Protocol" (p. 101)
• When internal (INT) triggering is selected, data is exported only when a TRIG signal is applied.
• Executing a
:READ?
query command exports duplicate measured values.
• For other queries, be careful to avoid overlapping query response timing with auto-exporting measured val-
ues.
• The data format for measured values can be selected as ASCII (default) or BINARY. Transfer time is mini-
mized when BINARY is selected.
See:
":SYSTem:FORMat <ASCii/ BINary>" (p. 137)
1
Open the Basic Settings screen.
2
Open the Data Memory Settings screen.
3
Enable or disable auto-exporting (DATA OUT)
4
Return to the Measurement screen.
The Basic Settings screen appears.
The Data Settings screen appears.
[DATA SETTINGS]
1
Selection
2
Selection
1
2
Disable auto-exporting (default)
Enable auto-exporting
The confirmation screen appears.
Return to the setting screen.
Save setting and return to
previous screen.
Discard setting and return to
previous screen.
Содержание RM3542
Страница 2: ......
Страница 6: ...Contents iv ...
Страница 16: ...Operating Precautions 10 ...
Страница 26: ...1 3 Screen Organization 20 ...
Страница 32: ...2 3 Turning the Power On and Off 26 ...
Страница 46: ...3 8 Confirming Faulty Measurements 40 ...
Страница 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Страница 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Страница 90: ...7 3 Printing 84 ...
Страница 172: ...9 10 Device Compliance Statement 166 ...
Страница 190: ...11 4 Disposing of the Instrument 184 ...
Страница 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Страница 214: ...Index Index 4 ...
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......