Appendix 9 Zero Adjustment
A
18
To perform zero adjustment using a probe
When you actually perform zero adjustment using a probe, you may unexpectedly
make the connection shown in Table 1 (b). Therefore, when performing zero
adjustment, you need to pay sufficient attention to the connection state of each
terminal.
Here, 9287-10 Clip Type Lead is used as an example for the connection explanation.
Table 2 shows the connection state of the tip of the lead and equivalent circuit in the
respective correct and wrong connections. Table 1 (a) indicates the correct
connection method, resulting in 0 V between H
POT
and L
POT
. However, Table 1 (b) is
the wrong connection method, so that 0 V is not obtained between H
POT
and L
POT
.
Table 2: Clip type lead connection methods used during zero adjustment
Connection method
Correct
Wrong
Tip of lead
Equivalent circuit
Deformed equivalent
circuit
As connection
method for zero
adjustment
Correct
Wrong
Red
Black
Red
Black
Содержание RM3542
Страница 2: ......
Страница 6: ...Contents iv ...
Страница 16: ...Operating Precautions 10 ...
Страница 26: ...1 3 Screen Organization 20 ...
Страница 32: ...2 3 Turning the Power On and Off 26 ...
Страница 46: ...3 8 Confirming Faulty Measurements 40 ...
Страница 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Страница 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Страница 90: ...7 3 Printing 84 ...
Страница 172: ...9 10 Device Compliance Statement 166 ...
Страница 190: ...11 4 Disposing of the Instrument 184 ...
Страница 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Страница 214: ...Index Index 4 ...
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......