9.7 Message Reference
138
(7) Triggering
Relationship Between Trigger Source and Continuous Measurement Operation
Operation depends on the continuous measurement setting (
:INITIATE:CONTINUOUS
,
the trigger source setting (
:TRIGGER:SOURCE
,
), as follows.
See:
"9.8 Data Exporting Methods" (p. 148)
Measurement Flow
Continuous Measurement Command-Specific Settings
:INITIATE:CONTINUOUS ON
:INITIATE:CONTINUOUS OFF
Trigger
Source
:TRIGGER:SOURCE IMM
Free-Run state.
Measurement continues automatically.
Trigger by
:INITIATE
(or
:READ?
) com-
mand.
:TRIGGER:SOURCE EXT
Trigger by TRIG signal,
F4
[MANU] key or
TRG
command. After measurement, enters
the trigger wait state.
Issue
:INITIATE
(or
:READ?
) command
to wait for trigger. Trigger by TRIG signal.
:INITIATE:CONTINUOUS OFF
Can only be set by Remote command.
If this has been set to OFF when operation is returned to the Local state or power is turned off, the following state occurs
when power is turned back on.
:INITIATE:CONTINUOUS ON
("To cancel the Remote state (enter the local state)" (p. 104)
Exporting measured values: "9.8 Data Exporting Methods" (p. 148)
Measurement
Calculation
Measured Value Output
Trigger Delay
Idle State
Measurement
Calculation
Measured Value Output
Trigger Delay
:INITIATE:IMMEDIATE
or
:READ?
Trigger Wait State
Measurement
Calculation
Measured Value Output
Trigger Delay
TRIG signal, F4 [MANU]
key,
TRG
Trigger Wait State
Measurement
Calculation
Measured Value Output
Trigger Delay
Idle State
TRIG signal
:INITIATE:IMMEDIATE
or
:READ?
Содержание RM3542
Страница 2: ......
Страница 6: ...Contents iv ...
Страница 16: ...Operating Precautions 10 ...
Страница 26: ...1 3 Screen Organization 20 ...
Страница 32: ...2 3 Turning the Power On and Off 26 ...
Страница 46: ...3 8 Confirming Faulty Measurements 40 ...
Страница 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Страница 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Страница 90: ...7 3 Printing 84 ...
Страница 172: ...9 10 Device Compliance Statement 166 ...
Страница 190: ...11 4 Disposing of the Instrument 184 ...
Страница 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Страница 214: ...Index Index 4 ...
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......