3.2 Measurement Object Types
28
The instrument provides two measurement methods: resistance measurement, and low-power
resistance measurement. Select the appropriate measurement method for the type of component to
be measured. For general-purpose resistor measurements, use the factory defaults. The power
applied to the DUT = Resistance Value
×
(Measurement Current)
2
.
See:
"(6) DUT Becomes Warm" (p. A6)
(Example)If the resistance to be measured is 100
The 1000
and higher ranges (with LOW POWER: OFF) are not usable for inductor measurements.
3.2
Measurement Object Types
(Measurement Current)
(Measurement Method)
10 mA
100
×
0.01
2
= 10 mW
Normal Resistance Measurement, 100
Range
1 mA
100
×
0.001
2
= 100
W
Low-Power Resistance Measurement, 100
Range
Measurement range: 0.0000 m
(100 m
range) to
120.0000 M
(10 ranges)
Normal Resistance Measurement
General-purpose resistors
Measurement range: 0.000 m
(1000m
range) to
1200.000
(4 ranges)
Low-Power Resistance Measurement
Hard-to-measure components such
as ferrite bead or layered inductors,
or other elements sensitive to mea-
surement current
LP
LP appears at the top of the screen.
1
Open the Basic Settings screen.
2
Open the System screen.
3
Select the low-power mode, as needed.
4
Return to the Measurement screen.
The Basic Settings screen appears.
1
The System screen appears.
[SYSTEM]
Selection
2
Selection
1
2
Normal resistance measurement (default)
Low-power resistance measurement
Return to the setting screen.
Save setting and return to
previous screen.
Discard setting and return to
previous screen.
The confirmation screen appears.
Содержание RM3542
Страница 2: ......
Страница 6: ...Contents iv ...
Страница 16: ...Operating Precautions 10 ...
Страница 26: ...1 3 Screen Organization 20 ...
Страница 32: ...2 3 Turning the Power On and Off 26 ...
Страница 46: ...3 8 Confirming Faulty Measurements 40 ...
Страница 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Страница 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Страница 90: ...7 3 Printing 84 ...
Страница 172: ...9 10 Device Compliance Statement 166 ...
Страница 190: ...11 4 Disposing of the Instrument 184 ...
Страница 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Страница 214: ...Index Index 4 ...
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......