4.6 Detecting Measurement Voltage Faults (Voltage Level Monitor Function)
49
4
When a measurement voltage fault occurs due to probe chattering, the
C.E. Volt
error message
appears on the measurement screen and an
ERR
signal is output.
The
C.E. Volt
error may also appear when external noise is strong.
Check the following if errors occur frequently:
• Check for probe deterioration.
• Provide additional noise suppression. "Appendix 3 Unstable Measurement Values" (p. A3)
• Set the voltage level monitor to Loose, or OFF (disable).
4.6
Detecting Measurement Voltage Faults
(Voltage Level Monitor Function)
1
Open the Basic Settings screen.
2
Open the Measurement Settings Screen.
3
Enable or disable the function.
4
(When enabled (ON selected))
Select the voltage level monitor threshold.
5
Return to the Measurement screen.
The Basic Settings screen appears.
The Measurement Settings Screen
appears.
[MEAS SETTINGS]
1
Selection
2
Selection
1
2
Disables the function (go to step 5)
Enables the function (default)
The setting is specific to the selected range (p. 41)
Selection
1
2
Loose *
Normal*
Severe
*. Default setting: Loose is
the default for the 100 M
range, and Normal for all
other ranges.
The confirmation screen appears.
Return to the setting screen.
Save setting and return to
previous screen.
Discard setting and return to
previous screen.
Содержание RM3542
Страница 2: ......
Страница 6: ...Contents iv ...
Страница 16: ...Operating Precautions 10 ...
Страница 26: ...1 3 Screen Organization 20 ...
Страница 32: ...2 3 Turning the Power On and Off 26 ...
Страница 46: ...3 8 Confirming Faulty Measurements 40 ...
Страница 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Страница 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Страница 90: ...7 3 Printing 84 ...
Страница 172: ...9 10 Device Compliance Statement 166 ...
Страница 190: ...11 4 Disposing of the Instrument 184 ...
Страница 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Страница 214: ...Index Index 4 ...
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......