7.1 Connecting the Printer
79
7
Before connecting the printer
Compatible printer
The requirements for a printer to be connected to the instrument are as follows.
Confirm compatibility and make the appropriate settings on the printer before connecting it to
the instrument.
See:
"7.2 Instrument Settings" (p. 81)
• Interface ................................ RS-232C
• Characters per line ................ At least 45
• Communication speed........... 9600 bps
• Data bits ................................ 8
• Parity ..................................... none
• Stop bits................................. 1
• Flow control ........................... none
Printing
Chapter 7
Printing (p. 82)
• Measurement values and
comparator decisions
• Statistical calculation
results
Connecting the printer to
the instrument
Make instrument
settings (p. 81)
Make printer
settings
7.1
Connecting the Printer
Because electric shock and instrument damage hazards are present, always follow
the steps below when connecting the printer.
•
Always turn off the instrument and the printer before connecting.
•
A serious hazard can occur if a wire becomes dislocated and contacts another con-
ductor during operation. Make certain connections are secure.
• As much as possible, avoid printing in hot and humid environments. Otherwise, printer life
may be severely shortened.
• Use only compatible recording paper in the printer. Using non-specified paper may not only
result in faulty printing, but printing may become impossible.
• If the recording paper is skewed on the roller, paper jams may result.
Содержание RM3542
Страница 2: ......
Страница 6: ...Contents iv ...
Страница 16: ...Operating Precautions 10 ...
Страница 26: ...1 3 Screen Organization 20 ...
Страница 32: ...2 3 Turning the Power On and Off 26 ...
Страница 46: ...3 8 Confirming Faulty Measurements 40 ...
Страница 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Страница 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Страница 90: ...7 3 Printing 84 ...
Страница 172: ...9 10 Device Compliance Statement 166 ...
Страница 190: ...11 4 Disposing of the Instrument 184 ...
Страница 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Страница 214: ...Index Index 4 ...
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......