9.7 Message Reference
141
9
Time to receive measured values is different for the
:FETCh?
and
:READ?
commands.
See:
"9.8 Data Exporting Methods" (p. 148)
Reading the Most Recent Measurement
Syntax
Query
:FETCh?
Description
Reads the most recent measurement. No trigger occurs.
See:
"Measurement Value Formats" (p. 140)
Example
:FETC?
1023.579E-3
Note
Binary data is returned when the Binary sending format is selected.
Measuring (Awaiting Triggers and Reading Measurements)
Syntax
Query
:READ?
Description
Switches from the Idle State to the Trigger Wait State, then reads the next measured value. With
auto-ranging enabled, the most suitable range is selected before measurement.
See:
"Measurement Value Formats" (p. 140)
Error
•
This command causes an execution error if issued during the Continuous Measurement state
(after :INITIATE:CONTINUOUS ON).
•
This command causes an execution error if issued during the Trigger Wait State.
Note
•
The next command does not execute until measurement is finished.
•
When using an external trigger source (EXT) with auto-sending of measured values (Auto-
Memory function) enabled, returned measured values will be duplicated. Therefore, disable the
Auto-Memory function when triggering externally.
See:
"6.4 Auto-Exporting Measured Values (at End of Measurement) (Data Output Function)" (p.
77)
•
Binary data is returned when the Binary sending format is selected.
Trigger Source Operation
IMMediate
Triggers and reads measured value.
EXTernal
Triggers by TRIG signal (EXT I/O) input, and continuously reads
measured values.
Содержание RM3542
Страница 2: ......
Страница 6: ...Contents iv ...
Страница 16: ...Operating Precautions 10 ...
Страница 26: ...1 3 Screen Organization 20 ...
Страница 32: ...2 3 Turning the Power On and Off 26 ...
Страница 46: ...3 8 Confirming Faulty Measurements 40 ...
Страница 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Страница 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Страница 90: ...7 3 Printing 84 ...
Страница 172: ...9 10 Device Compliance Statement 166 ...
Страница 190: ...11 4 Disposing of the Instrument 184 ...
Страница 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Страница 214: ...Index Index 4 ...
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......