4.11 Maintaining Measurement Precision (Self-Calibration)
57
4
To maintain measurement precision, the instrument self-calibrates every ten minutes to compensate
for internal circuit offset voltage and gain drift. This function cannot be disabled.
During self-calibration, the subsequent measurement is delayed for about 6PLC + 10 ms (PLC =
Power Line Cycles) for internal circuit compensation.
Self-Calibration Timing
Within 130 ms at 50 Hz, or 110 ms at 60 Hz
• When the timing of self-calibration overlaps with a measurement, self-calibration is postponed until the end
of measurement.
• When a trigger signal is applied during self-calibration, the start of the triggered measurement is postponed
until self-calibration is finished.
• Self-calibration executes automatically after changing comparator or measurement speed settings.
• During self-calibration, measurement current and the Contact Improver current are inhibited.
This function automatically compensates for offset voltage resulting from thermal emf or internal
instrument bias. (OVC: Offset Voltage Compensation)
See:
"Appendix 2 Effect of Thermal emf" (p. A2)
The following value is known to be a true resistance value from R
P
(>0), the value measured with
current flowing in the positive direction, and R
N
(<0), the value measured with current flowing in the
negative direction.
Offset voltage compensation is automatically enabled in the following conditions, and cannot be modified or
disabled:
• When a range from 100 m
to 10
is selected.
• When low-power resistance measurement is enabled (LOW POWER: ON).
When the test object is inductive, some delay (DELAY2) is required (p. 42) to allow adequate current flow
before starting measurement.
4.11 Maintaining Measurement Precision
(Self-Calibration)
4.12 Compensating for Thermal EMF Offset
(Offset Voltage Compensation - OVC)
R
p
R
N
–
2
----------------------
Содержание RM3542
Страница 2: ......
Страница 6: ...Contents iv ...
Страница 16: ...Operating Precautions 10 ...
Страница 26: ...1 3 Screen Organization 20 ...
Страница 32: ...2 3 Turning the Power On and Off 26 ...
Страница 46: ...3 8 Confirming Faulty Measurements 40 ...
Страница 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Страница 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Страница 90: ...7 3 Printing 84 ...
Страница 172: ...9 10 Device Compliance Statement 166 ...
Страница 190: ...11 4 Disposing of the Instrument 184 ...
Страница 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Страница 214: ...Index Index 4 ...
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......