9.7 Message Reference
136
Compare and Query Measurement Settings on Two Instruments
(Settings Monitor Function)
Set and Query Settings Comparison
Syntax
Command
:SYSTem:SETMonitor
<1/ 0/ ON/ OFF>
Query
Response
:SYSTem:SETMonitor?
<ON/ OFF>
Description
When using two instruments, this function checks that comparator and other settings are the same
on both.
See:
"4.9 Comparing the Measurement Settings of Two Instruments (Settings Monitor Function)"
(p. 53)
Example
:SYST:SETM ON
Set and Query Settings Monitor Order (1
st
and 2
nd
Stages)
Syntax
Command
:SYSTem:SETMonitor:ORDer
<1/ 2>
Query
Response
:SYSTem:SETMonitor:ORDer?
<1/ 2 (NR1)>
Example
:SYST:SETM:ORD 1
Set and Query Settings Monitor Tolerance
Syntax
Command
:SYSTem:SETMonitor:ALLowance
<0 to 9.999 (%)>
Query
Response
:SYSTem:SETMonitor:ALLowance?
<0 to 9.999
(NR1)>[%]
Example
:SYST:SETM:ALL 0.5
Return to Local Control
Syntax
Command
:SYSTem:LOCal
Description
Disables communications remote control and re-enables local control. The panel keys are re-
enabled.
Example
:SYST:LOC
Set and Query Measurement-Synchronized Data Output
Syntax
Command
:SYSTem:DATAout
<1/ 0/ ON/ OFF>
Query
Response
:SYSTem:DATAout?
<ON/ OFF>
Description
ON: The measured value is automatically sent each time an externally triggered measurement is
finished. During internal triggering, measured values are sent automatically whenever the TRIG
signal is applied.
OFF: Measured values are not automatically sent.
This command is not applicable to the GP-IB Interface.
Содержание RM3542
Страница 2: ......
Страница 6: ...Contents iv ...
Страница 16: ...Operating Precautions 10 ...
Страница 26: ...1 3 Screen Organization 20 ...
Страница 32: ...2 3 Turning the Power On and Off 26 ...
Страница 46: ...3 8 Confirming Faulty Measurements 40 ...
Страница 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Страница 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Страница 90: ...7 3 Printing 84 ...
Страница 172: ...9 10 Device Compliance Statement 166 ...
Страница 190: ...11 4 Disposing of the Instrument 184 ...
Страница 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Страница 214: ...Index Index 4 ...
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......