9.7 Message Reference
144
(10)Measurement Settings
Set and Query Delay 2 (after applying measurement current)
Syntax
Command
:RESistance:DELay2
<Range>
,
<Delay 2 time>
Query
Response
:RESistance:DELay2?
<Range>
<Delay 2 time>
Command
:LPResistance:DELay2
<Range>
,
<Delay 2 time>
Query
Response
:LPResistance:DELay2?
<Range>
<Delay 2 time>
<Range>
=
(:RESistance) RNG100MIL/ RNG1000MIL/ RNG10/ RNG100/ RNG1000/ RNG10K/
RNG100K/ RNG1000K/ RNG10MEG/ RNG100MEG
(:LPResistance) RNG1000MIL/ RNG10/ RNG100/ RNG1000
<Delay 2 time>
= 0 to 0.1 (seconds)
Example
:RES:DEL2 RNG1000MIL,0.003
:RES:DEL2? RNG10
0.005
Set and Query Integration Time (no. of power line cycles)
Syntax
Command
:RESistance:NPLCycles
<Range>
,
<Speed>
,
<Integration time
(NPLC)>
Query
Response
:RESistance:NPLCycles?
<Range>
,
<Speed>
<Integration time (NPLC)>
Command
:LPResistance:NPLCycles
<Range>
,
<Speed>
,
<Integration time
(NPLC)>
Query
Response
:LPResistance:NPLCycles?
<Range>
,
<Speed>
<Integration time (NPLC)>
<Range>
=
(:RESistance) RNG100MIL/ RNG1000MIL/ RNG10/ RNG100/ RNG1000/ RNG10K/
RNG100K/ RNG1000K/ RNG10MEG/ RNG100MEG
(:LPResistance) RNG1000MIL/ RNG10/ RNG100/ RNG1000
<Speed>
= SLOW/ MEDium/ FAST
<Integration time (NPLC)>
= 0.01 to 6
Description
Sets the integration time as an integer multiple of power line cycles (NPLC).
Note
•
For high-resistance measurements, hum noise is suppressed and stable measured values are
acquired by setting an integer multiple of power line cycles.
•
When converting to seconds, values over 100 ms are set to 100 ms.
•
When querying the integration time, the instrument’s internal variable is converted and output.
So in some cases, the returned value may not match the set value.
Содержание RM3542
Страница 2: ......
Страница 6: ...Contents iv ...
Страница 16: ...Operating Precautions 10 ...
Страница 26: ...1 3 Screen Organization 20 ...
Страница 32: ...2 3 Turning the Power On and Off 26 ...
Страница 46: ...3 8 Confirming Faulty Measurements 40 ...
Страница 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Страница 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Страница 90: ...7 3 Printing 84 ...
Страница 172: ...9 10 Device Compliance Statement 166 ...
Страница 190: ...11 4 Disposing of the Instrument 184 ...
Страница 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Страница 214: ...Index Index 4 ...
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......