9.7 Message Reference
139
9
Continuous Measurement Setting
Syntax
Command
:INITiate:CONTinuous
<1/ 0/ ON/ OFF>
Query
Response
:INITiate:CONTinuous?
<ON/ OFF>
<ON>
= Continuous Measurement Enabled
<OFF>
= Continuous Measurement Disabled
Description
•
Continuous Measurement Enabled:
After measurement, enters the Trigger Wait State. When the trigger source setting is IMMediate,
the next trigger occurs immediately (the Free-Run State).
•
Continuous Measurement Disabled:
After measurement, enters the Idle State instead of the Trigger Wait State.
•
Triggering is ignored in the Idle State. Executing :INITiate[:IMMediate] enables the Trigger Wait
State.
•
Continuous measurement is enabled upon exit from the Remote State.
Example
:INIT:CONT OFF
:INIT:CONT?
ON
Trigger Wait Setting
Syntax
Command
:INITiate[:IMMediate]
Description
Switches triggering from the Idle State to the Trigger Wait State.
Example
Disable continuous measurement, and read one value for each trigger event.
Sending
:TRIG:SOUR IMM
..Trigger immediately when entering Trigger Wait State
:INIT:CONT OFF
..Disables continuous measurement
:INIT
..........................Enable Trigger Wait Trigger immediately upon :TRIG:SOUR IMM
:FETC?
.......................Fetch measured value
Reading
2.3
.............Measured value is 2.16414 k
Error
An execution error occurs when continuous measurement is enabled.
(
:INITIATE:CONTINUOUS ON
).
Note
•
When the trigger source is IMMediate, triggering occurs immediately before entering the Idle
State.
•
When the trigger source is EXTernal, the Trigger Wait State is enabled to wait for an external
trigger, and when a trigger occurs, one measurement is taken before entering the Idle State.
Trigger Source Setting
Syntax
Command
:TRIGger:SOURce
<IMMediate/ EXTernal>
Query
Response
:TRIGger:SOURce?
<IMMEDIATE/ EXTERNAL>
<IMMEDIATE>
= Internal triggering
<
EXTERNAL>
= External triggering
Description
Trigger by pressing F4 [MANU] on the [MENU]-[TRG] selection screen, TRIG signal or
TRG
command.
Example
:TRIG:SOUR IMM
:TRIG:SOUR?
IMMEDIATE
Note
When external (EXT) triggering is enabled, the auto-memory mode is disabled (OFF).
Содержание RM3542
Страница 2: ......
Страница 6: ...Contents iv ...
Страница 16: ...Operating Precautions 10 ...
Страница 26: ...1 3 Screen Organization 20 ...
Страница 32: ...2 3 Turning the Power On and Off 26 ...
Страница 46: ...3 8 Confirming Faulty Measurements 40 ...
Страница 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Страница 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Страница 90: ...7 3 Printing 84 ...
Страница 172: ...9 10 Device Compliance Statement 166 ...
Страница 190: ...11 4 Disposing of the Instrument 184 ...
Страница 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Страница 214: ...Index Index 4 ...
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......