Chapter 10 Specifications
173
10
(24) Statistical Calculations
(25) Settings Monitor Function
(26) Retry Function
(27) Self-Test
(28) Trigger Source Setting
(29) Measurement Speed
(30) Key-Lock Functions
Operation
Statistical calculations are performed on measured values in memory.
Default state
OFF
Setting
ON/ OFF
Calculations
Total data count, Mean, Minimum value (sample no.), Maximum value (sample no.),
Standard deviation of sample, Population standard deviation, Process capability indices
Operation
Measurement settings of two instruments are compared, and if different, an alarm
sounds and TRIG input is inhibited.
However, the comparator thresholds of the 1
st
stage instrument can be smaller than
those of the 2
nd
.
Default state
OFF, 1st, 0.000%
Setting
Function
ON / OFF
Measurement stages1st / 2nd
Tolerance range 0.000% to 9.999%
Compared contents
Comparator threshold, measurement speed
Operation
The Retry function causes measurement to be automatically retried when a measure-
ment fault occurs due to probe chatter. If a measurement fault persists after the speci-
fied continuous retry interval, retrying is aborted and the EOM signal is output.
Default state
ON, 50 ms
Setting
Function ON/ OFF
Continuous retry interval 1 ms to 50 ms
At power-on
ROM/RAM check, non-volatile ROM checksum test
Default state
EXT (external)
Settings
INT (Internal)/ EXT (external)
Default state
FAST
Setting
FAST/ MED/ SLOW
Operation
Disables operation of unneeded keys.
Setting
(1) Disables all except the comparator setting and cancel keys
(2) Disables all except the cancel key
(3) All front panel keys are disabled when the KEY_LOCK signal is received.
Содержание RM3542
Страница 2: ......
Страница 6: ...Contents iv ...
Страница 16: ...Operating Precautions 10 ...
Страница 26: ...1 3 Screen Organization 20 ...
Страница 32: ...2 3 Turning the Power On and Off 26 ...
Страница 46: ...3 8 Confirming Faulty Measurements 40 ...
Страница 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Страница 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Страница 90: ...7 3 Printing 84 ...
Страница 172: ...9 10 Device Compliance Statement 166 ...
Страница 190: ...11 4 Disposing of the Instrument 184 ...
Страница 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Страница 214: ...Index Index 4 ...
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......