3.8 Confirming Faulty Measurements
38
When a measurement is not performed correctly, a measurement fault indicator appears and a
measurement fault signal is output at the ERR pin of the EXT I/O connector (except for out-of-range
detection).
The instrument detects measurement faults by the following four methods.
3.8
Confirming Faulty Measurements
Appears when the measured value is outside of the measurement or display range.
Check for a broken sample component.
The comparator result is Hi when +OvrRng is displayed, and Lo when -OvrRng is displayed.
No external measurement fault signal (ERR) is output.
Out-of-Range
The resistance between the H
POT
and H
CUR
probe contacts, and between the L
POT
and L
CUR
probe contacts, are measured and compared with specified contact fault values.
An error message appears when the measured value reaches or exceeds the specified contact fault
values.
If this error persists, probe wear or cable failure may be the cause.
If the error is not cleared by shorting the tips of a known-good measurement cable, the instrument
requires repair.
Contact Check Fault
This method monitors the stability of the voltage between H
POT
and L
POT
probe contacts.
An error message appears when voltage instability is detected due to chattering of the probe con-
tacts.
If this error persists, the probes may be degraded due to wear.
C.E. Volt may also be displayed when external noise is strong.
Voltage Level Monitor Fault
See:
"Out-of-Range Detection Function" (p. 39)
See:
"4.4 Checking for Poor or Improper Contact
(Contact Check Function)" (p. 46)
See:
"4.6 Detecting Measurement Voltage Faults
(Voltage Level Monitor Function)" (p. 49)
This display appears after changing measurement settings and before the next measurement is
performed.
- - - - - - -
Display Examples: Display Measurement State and Appearance with Open-Circuit Probe
Display Measurement State
Current Monitor Results
Normal (PASS)
Fault (FAIL)
Contact Check Results
Voltage Level Monitor
Results
Normal
(PASS)
Display: Measured Value
COMP indicator: According to the
measured value
Display:
+OvrRng/ -OvrRng
COMP indicator:
Hi/ Lo
(when connection to the measure-
ment object is broken)
Fault
(FAIL)
Display:
C.E. Hi/ C.E. Lo/ C.E. Volt
COMP indicator: No decision
EXT I/O: ERR signal output
Display:
C.E. Hi/ C.E. Lo/ C.E. Volt
COMP indicator: No decision
EXT I/O: ERR signal output
+OvrRng
-OvrRng
C.E. Hi
C.E. Lo
C.E. Volt
Display
Display
Display
This method monitors the regulated measurement current for normal flow through the DUT.
An error is detected mainly when a measurement fault occurs due to an open-circuit DUT or between the H
CUR
and
L
CUR
probe's poor contacts.
The error display depends on the contact check and voltage level monitor states (see the table below).
Current Monitor Fault
See:
"Current Monitor Function" (p. 39)
The measurement fault display differs according to detection order
and settings.
Содержание RM3542
Страница 2: ......
Страница 6: ...Contents iv ...
Страница 16: ...Operating Precautions 10 ...
Страница 26: ...1 3 Screen Organization 20 ...
Страница 32: ...2 3 Turning the Power On and Off 26 ...
Страница 46: ...3 8 Confirming Faulty Measurements 40 ...
Страница 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Страница 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Страница 90: ...7 3 Printing 84 ...
Страница 172: ...9 10 Device Compliance Statement 166 ...
Страница 190: ...11 4 Disposing of the Instrument 184 ...
Страница 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Страница 214: ...Index Index 4 ...
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......