2.2 Connecting Measurement Cables and Test Fixtures
24
Recommended Measurement Cable Specifications
Example: JIS standard 3C-2V, 1.5D-2V MIL standard RG-58A/U
* Teflon is a registered trademark of E. I. du Pont de Nemours and Company.
Making Your Own Measurement Cables
Conductor resistance
500 m
/m or less
Capacitance
150 pF/m or less
Cable dielectric material
Polyethylene (PE), Teflon* (TFE), polyethylene foam (PEF)
Insulation resistance at least 10 G
Connector insulating material
Teflon* (TFE), polybutylene terephtalate (PBT)
Insulation resistance at least 10 G
Length
2 m or less
Wiring Diagram
Before Wiring
• Twist together the H
POT
and L
POT
wires, and the H
CUR
and L
CUR
wires. If not twisted together, mea-
sured values may be unstable and errors occur when measuring with low-power resistance, or low resis-
tance values.
• Refer to the block diagram (p. 14) for internal circuit details.
• Probes and measurement objects should be shielded at BNC or GUARD jack potential.
• Measurement cable length should not exceed 2 m (with conductor resistance 500 m
/m or less).
Long cables are more susceptible to noise, and measured values may be unstable.
• Extensions should maintain the four-terminal structure. If converted to a two-terminal circuit in the
wiring, correct measurement may not be possible due to the effects of wiring and contact resis-
tance.
• Cables and measurement objects should be shielded.
• After extending measurement cables, confirm operation and accuracy ("Measurement Specifica-
• If cutting the ends off of optional measurement cables, make sure that the shield does not touch
the center conductor of the H
CUR
, H
POT
, L
POT
and L
CUR
cables. Correct measurement is not pos-
sible with a shorted cable.
Extending Measurement Cables
Observe the following when extending measurement cables:
Содержание RM3542
Страница 2: ......
Страница 6: ...Contents iv ...
Страница 16: ...Operating Precautions 10 ...
Страница 26: ...1 3 Screen Organization 20 ...
Страница 32: ...2 3 Turning the Power On and Off 26 ...
Страница 46: ...3 8 Confirming Faulty Measurements 40 ...
Страница 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Страница 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Страница 90: ...7 3 Printing 84 ...
Страница 172: ...9 10 Device Compliance Statement 166 ...
Страница 190: ...11 4 Disposing of the Instrument 184 ...
Страница 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Страница 214: ...Index Index 4 ...
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......