Appendix 1 Four-Terminal (Voltage-Drop) Method
A
1
Appendix
The Four-Terminal method is essential for measuring very small resistance values.
With two-terminal measurements (Fig. 1), the resistance of the test leads is included in the measured resis-
tance, resulting in measurement errors.
The four-terminal method (Fig. 2) consists of current source terminals (H
CUR
, L
CUR
) to provide constant cur-
rent, and voltage detection terminals (H
POT
, L
POT
) to detect voltage drop.
Because of the high input impedance of the voltmeter, measurement requires practically no current flow
through the leads connecting the voltage detection terminals to the test object, practically eliminating the
effects of lead and contact resistance on the measurement.
Appendix
Appendix 1
Four-Terminal (Voltage-Drop)
Method
Two-Terminal Measurement Method
Four-Terminal Measurement Method
Measurement current I flows through test object
resistance R
0
as well as lead resistances r
1
and
r
2
.
The voltage to be measured is obtained by E = I
(r
1
+ R
0
+ r
2
), which includes lead resistances r
1
and r
2
.
Current I flows from r
2
through DUT resistance
R
0
and through r
1
. The high input impedance of
the voltmeter allows only negligible current flow
through r
3
and r
4
.
So the voltage drop across r
3
and r
4
is practically
nil, and voltage E across the measurement
terminals and voltage E
0
across test object
resistance R
0
are essentially equal, allowing test
object resistance to be measured without being
affected by r
1
to r
4
.
Figure 1.
r
1
r
2
E
I
Resistance R
0
Voltmeter
Constant current source
Figure 2.
r
1
r
2
E
I
Resistance R
0
Voltmeter
Constant current source
r
3
r
4
E
0
Содержание RM3542
Страница 2: ......
Страница 6: ...Contents iv ...
Страница 16: ...Operating Precautions 10 ...
Страница 26: ...1 3 Screen Organization 20 ...
Страница 32: ...2 3 Turning the Power On and Off 26 ...
Страница 46: ...3 8 Confirming Faulty Measurements 40 ...
Страница 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Страница 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Страница 90: ...7 3 Printing 84 ...
Страница 172: ...9 10 Device Compliance Statement 166 ...
Страница 190: ...11 4 Disposing of the Instrument 184 ...
Страница 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Страница 214: ...Index Index 4 ...
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......