Appendix 6 Calibration
A
11
Appendix
Calibration Conditions
Calibration equipment
Please use the following for calibration equipment.
See:
" Connection Methods" (p. A12)
Resistance measurement equipment
• Fluke 5700A (10
or more) or equivalent
• Alpha Electronics CSR-100 m
or equivalent
• Alpha Electronics CSR-1
or equivalent
If the Fluke 5700A cannot be used, please use the following equipment.
When using the YOKOGAWA 2792 to calibration
Use the Four-terminal Lead from Hioki. Note that connection cannot be made with the Clip
Type Lead.
Appendix 6
Calibration
• Ambient temperature and humidity 23±5°C, 80%RH or less
• Warm-up time
30 minutes
• Power supply
100 to 240 V±10%, 50/60 Hz
• Measurement speed
SLOW
• Integration time
Default setting
• Contact Improver
OFF (Required when using the Fluke 5700A)
• Measurement Current
CONT (Required when using the Fluke 5700A)
• Probe short-circuit detection
OFF
• 0ADJ
OFF (necessary)
• External magnetic field
Environment close to the Earth’s magnetic field
• Voltage level monitor
OFF (Required when using the Fluke 5700A)
• Alpha Electronics CSR-100 (10
)
• Alpha Electronics CSR-105 (1 M
)
• Alpha Electronics CSR-101 (100
)
• Alpha Electronics CSR-106 (10 M
)
• Alpha Electronics CSR-102 (1 k
)
• Alpha Electronics CSR-107 (100 M
)
• Alpha Electronics CSR-103 (10 k
)
• Alpha Electronics CSR-104 (100 k
)
OK
Four-terminal Lead
Clip Type Lead
Содержание RM3542
Страница 2: ......
Страница 6: ...Contents iv ...
Страница 16: ...Operating Precautions 10 ...
Страница 26: ...1 3 Screen Organization 20 ...
Страница 32: ...2 3 Turning the Power On and Off 26 ...
Страница 46: ...3 8 Confirming Faulty Measurements 40 ...
Страница 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Страница 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Страница 90: ...7 3 Printing 84 ...
Страница 172: ...9 10 Device Compliance Statement 166 ...
Страница 190: ...11 4 Disposing of the Instrument 184 ...
Страница 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Страница 214: ...Index Index 4 ...
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......