Using Analog Output
7.6 Using Analog Output
When analog output is used, a DC voltage that is proportional to the resistance is output from the
analog output terminal on the back of the instrument.
Analog output is updated at the same time as the instrument displays the measured value. When
the test is ended, the analog output is held with the final voltage output.
A voltage is output according to the resistance range as shown in the following table.
4 V is output when the maximum value of each resistance range is displayed.
Resistance Range
Resistance value range
Output voltage (DC)
2 M
Ω
0.000 M
Ω
to 9.999 M
Ω
0 V to 4 V
20 M
Ω
0.00 M
Ω
to 99.99 M
Ω
0 V to 4 V
200 M
Ω
0.0 M
Ω
to 999.9 M
Ω
0 V to 4 V
2000 M
Ω
(100 V ≤ V ≤ 500 V)
0 M
Ω
to 9999 M
Ω
0 V to 4 V
Total resistance range
Over.F
4 V
Under.F
0 V
IMPORTANT
When the resistance falls below the resistance display range,
[Under. F]
is displayed on the
screen and 0 V is output from the analog output terminal.
See “Resistance measurement” (p. 180).