173
Communications Command Examples
Example of using a contact check
The following is an example setting and enabling the threshold value of the contact check.
Description
Sent to the BT5525 from
the controller
Example of the BT5525
response
(1) Set the judgment threshold
value of the contact check.
:CONTactcheck:CAPacitance:
THReshold 0.5E-9
–
(2) Enable the contact check
function.
:CONTactcheck ON
–
(3)
Test start
:TIMer 2;:STARt
–
(4) Acquire the measurement
status.
(Monitor status by
periodically sending
:STATe?
.)
:STATe?
1
:STATe?
1
…
:STATe?
2
:STATe?
0
(5)
Query the contact check
capacitance value.
:CONTactcheck:CAPacitance?
1.2E-09
(6)
Query the contact check
result.
:CONTactcheck:RESult?
PASS
(1)
Set the PASS judgment threshold value of the contact check to 500 pF.
The contact check threshold value can be set in the range between 0.1E-9 and 200E-9 (100 pF to
200 nF).
(2)
Enable the contact check function.
(3)
Perform the test for 2 seconds.
(4)
Query the test status and monitor until the test is complete.
(5)
After the test is complete, query the measured capacitance value of the contact check.
1.2 nF is returned.
(6)
Query the contact check result.
PASS is returned since the result is over the setting, 500 pF.
Communications Function
HIOKI BT5525A961-00
Содержание BT5525
Страница 2: ...HIOKI BT5525A961 00 ...
Страница 6: ...Contents 6 HIOKI BT5525A961 00 ...
Страница 48: ...48 Notifying the Judgment Result and Test Completion with a Beep HIOKI BT5525A961 00 ...
Страница 76: ...76 Initializing the Instrument Reset HIOKI BT5525A961 00 ...
Страница 82: ...82 Deleting Panel Data HIOKI BT5525A961 00 ...
Страница 200: ...200 License Information HIOKI BT5525A961 00 ...
Страница 215: ...www hioki com HIOKI BT5525A961 00 ...