197
Error
10.3 Error
Error
No.
Error description
Details
Solution
EXT
. I/O
error signal
Response of
measurement
status
Response to the error query
(
:SYSTem:ERRor?
)
Bit of the Status Byte Register (STB) to be set
Bit of the
Standard Event Status Register (SESR) to be
set
Clear
by
*CLS
command
0
No error
There is no error
.
0,"No Error"
-100
Command error
The character string of the communications command is invalid.
Check that the command character string is correct.
-100,"Command error"
ERR
CME
-200
Execution error
The communications command cannot be executed due to
specific conditions.
The command cannot be executed under certain conditions.
-200,"Execution error"
ERR
EXE
-220
Parameter error
The communications command parameters are out of the range or invalid.
Check the command parameter range or character string.
-220,"Parameter error"
ERR
EXE
-316
Overheat error
An output circuit overheat event was detected, and output was stopped.
Allow the instrument to
cool.
Output will be possible 1 s after output stops.
20
-316,
”
Overheat
error
”
ERR
DDE
-330
Self-test error
The result of the self-test (
*TST?
) was F
AIL
Repair is necessary
.
-330,"Self-test failed"
ERR
DDE
-360
RS-232C communications error
A break signal was detected during RS-232C communications.
-360,"Communication error"
ERR
DDE
-361
RS-232C parity error
A parity error occurs during RS- 232C communications.
Check that the controller data length is 8 bits with no parity
.
Implement a noise countermeasure, such as attaching a ferrite core to the RS-232C line.
-361,"Rs232c Parity error"
ERR
DDE
-362
RS-232C framing error
A framing error occurs during RS-232C communications.
Check that the controller data length is 8 bits with no parity
.
Implement a noise countermeasure, such as attaching a ferrite core to the RS-232C line.
-362,"Rs232c Framing error"
ERR
DDE
-363
RS-232C overrun error
An overrun error occurs during RS-232C communications.
Reduce the data volume to be sent from the controller at one time.
Change the RS-232C communications speed.
-363,"Rs232c overrun"
ERR
DDE
-380
ROM error
A ROM error has been detected in a self-test at the startup.
Repair is necessary
.
ON output
99
-380, "ROM error."
ERR
DDE
Not cleared
-381
RAM error
A RAM error has been detected in a self-test at the startup.
Repair is necessary
.
ON output
99
-381,"RAM error."
ERR
DDE
Not cleared
-382
FRAM error
A FRAM error has been detected in a self-test at the startup.
Repair is necessary
.
ON output
99
-382,"FRAM error."
ERR
DDE
Not cleared
Maintenance and Service
HIOKI BT5525A961-00
Содержание BT5525
Страница 2: ...HIOKI BT5525A961 00 ...
Страница 6: ...Contents 6 HIOKI BT5525A961 00 ...
Страница 48: ...48 Notifying the Judgment Result and Test Completion with a Beep HIOKI BT5525A961 00 ...
Страница 76: ...76 Initializing the Instrument Reset HIOKI BT5525A961 00 ...
Страница 82: ...82 Deleting Panel Data HIOKI BT5525A961 00 ...
Страница 200: ...200 License Information HIOKI BT5525A961 00 ...