37
Setting the Range (Auto/Manual)
Setting the test voltage in the Setting screen
1
Open the Setting screen.
[SET]
Select
2
Set
[RANGE]
to
[AUTO]
or
[HOLD]
.
Select
[AUTO]
Set the auto range
[HOLD]
Set the manual range
3
Select the range when the manual range is selected.
Select
[ ↓ ]
Lower the range
[ ↑ ]
Raise the range
[EXIT]
Return to the Measurement screen
IMPORTANT
It may take time for measured values to stabilize in the following circumstances:
• When using the 200 M
Ω
range with a set voltage of 25 V to 99 V
• When using the 2000 M
Ω
range with a set voltage of 100 V to 500 V
When the auto range is selected,
[AUTO]
is displayed in the Measurement screen.
See “Measurement screen” (p. 21).
When analog output is used, 0 V to 4 V is output from the analog output terminal according to each
resistance range.
See “7.6 Using Analog Output” (p. 97).
Basic Settings
HIOKI BT5525A961-00
Содержание BT5525