130
Communications Command Reference
8.9 Communications Command Reference
Command format example
11
Setting the power supply frequency
Syntax
Command
:SYSTem:LFRequency <Power supply frequency>
•
Power supply frequency {AUTO|50|60}
Sets the power supply frequency.
Description
Specifies the power supply frequency.
AUTO
The power supply frequency (50 Hz or 60 Hz) is automatically detected at
the time of startup.
50
The power supply frequency is set to 50 Hz.
60
The power supply frequency is set to 60 Hz.
Example
:SYSTem:LFRequency 60
The power supply frequency is set to 60 Hz.
2
2
33
44
55
11
< >
Characters in these symbols indicate a parameter required for sending a command.
[ ]
Characters in these brackets can be omitted.
{ }
Placing characters in these brackets indicates that one of the items in these brackets
needs to be selected. Items are separated with a vertical bar (|). Numerical values in these
brackets indicate a setting range.
22
Indicates command information.
33
Indicates the syntax of communications commands.
The command data or response message is described.
44
Describes the command.
55
Indicates an example of using the actual command.
The blank (0x20) in the query example is indicated as “
␣
”.
Command, query
Response
HIOKI BT5525A961-00
Содержание BT5525
Страница 2: ...HIOKI BT5525A961 00 ...
Страница 6: ...Contents 6 HIOKI BT5525A961 00 ...
Страница 48: ...48 Notifying the Judgment Result and Test Completion with a Beep HIOKI BT5525A961 00 ...
Страница 76: ...76 Initializing the Instrument Reset HIOKI BT5525A961 00 ...
Страница 82: ...82 Deleting Panel Data HIOKI BT5525A961 00 ...
Страница 200: ...200 License Information HIOKI BT5525A961 00 ...
Страница 215: ...www hioki com HIOKI BT5525A961 00 ...