Setting the Sampling Time
3.3 Setting the Sampling Time
Set the sampling time in the range between 1 PLC and 100 PLCs.
1 PLC is equivalent to one cycle of a power supply frequency.
50 Hz
1 PLC = 1/50 = 20 ms
60 Hz
1 PLC = 1/60 = 16.7 ms
When the measured value is difficult to read due to a fluctuation, increase the sample time value.
1
Open the Setting screen.
[SET]
Select
2
Set the sampling time.
Select
[EDIT]
Edit
•Change the value (up/down keys)
•Change the cursor position
(left/right keys)
Confirm
[EXIT]
Return to the Measurement screen
IMPORTANT
If the test time is shorter than the sampling time, the measured value is not displayed. Set a
longer test time than the sampling time.