65
Limiting the Current to Be Applied to the Device Under Test (DUT)
Setting the current limit value manually
1
Open the Setting screen.
[SET]
Select
2
Set
[CURRENT LIMIT]
to
[MANUAL]
.
Select
[MANUAL]
Select
3
Set
[CURRENT LIMIT]
to the edit mode.
Select
[EDIT]
Edit
•
Change the value (up/down keys)
•
Change the cursor position
(left/right keys)
Confirm
[EXIT]
Return to the Measurement screen
IMPORTANT
•
If the set current limit value is from 5.1 mA to 50.0 mA, the current will be limited to 5 mA after
the output voltage reaches the set voltage.
•
If a capacitive load of approximately 50 μF or more is connected while the current limit value
of 5.1 mA or more, constraints related to the output generator will result in an error, making
measurement impossible.
See “9.2 Input, Output, and Measurement Specifications” (p. 180)
Various Functions