92
Internal Circuit Configuration
PNP setting
10
11
PLC, etc.
Output
Common
ISO_5 V
8
2 k
Ω
1 k
Ω
EXT. I/O
MODE
selector
START
1
STOP
20
SYSTEM_ERR
UPPER FAIL
9
ISO_COM
27
ISO_COM
PNP
10
Ω
PLC, etc.
Input
Common
Zener voltage 30 V
BT5525
Internal isolated power common
(It is isolated from the protective ground
of the instrument.)
Internal isolated
power supply
External power supply
Do not connect an external power supply to pin 8.
• Use ISO_COM for both input and output signals as a common terminal.
•
When power is supplied from an external instrument, supply the power to the external power
supply unit shown above.
HIOKI BT5525A961-00