209
LOW Terminal Side Test Lead Fabrication
Processing the tip of the LOW terminal side test lead
Completion image
Stripped core wire length = X mm
Heat shrinkable tube
1
Strip the coating using a knife, etc.
Braided conductor for the shielding wire
Sheath
X + 30 mm or more
(recommended value)
2
Fold back the braided conductor for the shielded cable.
Insulator
Braided conductor for the shielding wire
3
Remove the insulator using a knife, etc.
Braided conductor for the guard wire
Insulator
X +10
±
4 mm
(recommended value)
4
Fold back the braided conductor for the guard wire.
Secure 6 mm or more for Y to prevent short circuit.
(Adjust the length by cutting the braided conductor for
the guard wire.)
Y
Black conductive vinyl
Take care not to short the shielded cable and guard wire.
5
Remove the conductive vinyl to the base.
Appendix