Setting the Measurement Delay Time
3.4 Setting the Measurement Delay Time
Set the delay time that is allowed to elapse from the start of the test to the start of insulation
resistance measurement.
By setting a measurement delay time equal to interval during which measured values exhibit
instability immediately after the test voltage is applied, you can realize stable insulation resistance
measurement.
Set the measurement delay time in the range between 1 PLC and 100 PLCs.
1 PLC is equivalent to one cycle of a power supply frequency.
50 Hz
1 PLC = 1/50 = 20 ms
60 Hz
1 PLC = 1/60 = 16.7 ms
1
Open the Setting screen.
[SET]
Select
2
Set the measurement delay time.
Select
[EDIT]
Edit
•Change the value (up/down keys)
•Change the cursor position
(left/right keys)
Confirm
[EXIT]
Return to the Measurement screen