150
Communications Command Reference
Querying the BDD measured values saved in the memory
Syntax
Query
:BDD:MEMory?
[CRLF,{CCV|CVV|CVI}]
Response
<BDD measured value 1>,<BDD measured value 2>,
<BDD measured value 3>,
,<BDD measured value N>
Description
Queries all the BDD measured values saved in the memory.
If the optional BDD detection method is specified, the command returns the BDD
measured value detected using that method.
The BDD measured values are returned with comma delimiters. When
CRLF
is specified
as an option, they are returned with line break (
CRLF
) delimiters.
The number of BDD measured values, N, can be checked with the
:BDD:COUNt?
query.
An execution error occurs if the BDD measured values are not saved in the memory.
Response format
to
to
C C V
C V V
1 2 3 4
6
,
5
,
5 . 3 1
C
C
.
7
V
0 . 0 0 1
9 9 9 9 9 9 . 9 9 9
C V I
0 . 0 1
9 9 9 9 9 . 9 9
Time stamp (ms) (10 bytes)
Detection timing
and type (3 bytes)
V is detected during CC.
V is detected during CV.
I is detected during CV.
Voltage value (V) or current value (%)
(8 bytes)
Example
:BDD:MEMory?
16.640,CCV, 1.21, 33.280,CCV, 1.89
Two BDD measured values are saved in the memory and the result is as follows.
Time stamp
Detection timing and type
BDD measured value
16.640
CCV
1.21
33.280
CCV
1.89
Setting the BDD stop function
Syntax
Command
:BDD:STOP {ON|OFF}
Description
Sets whether to stop the test when the BDD count is 1 or more.
Example
:BDD:STOP ON
The test is set to stop when the BDD count is 1 or more.
Querying the BDD stop function
Syntax
Query
:BDD:STOP?
Response
(2 or 3 bytes +
CRLF)
{ON|OFF}
Description
Queries whether to stop the test when the BDD count is 1 or more.
Example
:BDD:STOP?
ON
The test is to stop when the BDD count is 1 or more.
HIOKI BT5525A961-00
Содержание BT5525
Страница 2: ...HIOKI BT5525A961 00 ...
Страница 6: ...Contents 6 HIOKI BT5525A961 00 ...
Страница 48: ...48 Notifying the Judgment Result and Test Completion with a Beep HIOKI BT5525A961 00 ...
Страница 76: ...76 Initializing the Instrument Reset HIOKI BT5525A961 00 ...
Страница 82: ...82 Deleting Panel Data HIOKI BT5525A961 00 ...
Страница 200: ...200 License Information HIOKI BT5525A961 00 ...
Страница 215: ...www hioki com HIOKI BT5525A961 00 ...