168
Communications Command Reference
Setting the auto data output function
Example
:SYSTem:COMMunicate:DATAout RS232C,TYPE1
:STARt
:STOP
␣␣␣␣
1 0.150 Mohm UFAIL
:STARt
:STOP
␣␣␣␣
2 2.154 Mohm PASS
The resistance value of the 1st measurement is 0.150 M
Ω
and the judgment is UFAIL.
The resistance value of the 2nd measurement is 2.154
Ω
and the judgment is PASS.
:SYSTem:COMMunicate:DATAout RS232C,TYPE2
:MEASure:VALid 255
:STARt
:STOP
␣␣␣␣
33, 0,0.153E+06, UFAIL,+2.01,+1.62918E-04,99,PASS
:STARt
:STOP
␣␣␣␣
33, 0,0.151E+06, UFAIL,+2.01,+1.65687E-04,99,PASS
Querying the auto data output function
Syntax
Query
:SYSTem:COMMunicate:DATAout?
Response
(8 to 12 bytes +
CRLF)
<Interface>,<Output type>
•
Interface {OFF|RS232C|USB|LAN}
Returns the interface setting to be used for automatic data output.
•
Output type {TYPE1|TYPE2}
Returns the data output format.
Description
Queries the interface to which data is automatically output and the data format.
Example
:SYSTem:COMMunicate:DATAout?
RS232C,TYPE1
The interface to which data is automatically output is RS-232C and the data format is TYPE1.
HIOKI BT5525A961-00