34
Setting the Test Voltage
3.1 Setting the Test Voltage
Set the test voltage in the range between 25 V and 500 V.
As the measurement range is determined based on the test voltage, any measurement range not
suitable for the setting is automatically changed. The test voltage can be set in the Measurement
screen or Setting screen. When the test voltage is set, the instrument will stop accepting operations
for about 1 second to allow the internal circuit’s voltage to stabilize. Please wait for the setting to
complete.
CAUTION
Set the test voltage correctly.
An inappropriate test voltage can damage the DUT.
Setting the test voltage in the Measurement screen
1
Set the test voltage to the edit mode.
[VOLT]
Select
2
Change and confirm the test voltage value.
•Change the value (up/down).
•Change the cursor position (left/right).
Select the preset voltage.
[25V]
[100V]
[200V]
[500V]
Confirm
[CANCEL]
Cancel
HIOKI BT5525A961-00