198
Error
Error
No.
Error description
Details
Solution
EXT
. I/O
error signal
Response of
measurement
status
Response to the error query
(
:SYSTem:ERRor?
)
Bit of the Status Byte Register (STB) to be set
Bit of the
Standard Event Status Register (SESR) to be
set
Clear
by
*CLS
command
-383
Internal temperature error
(instrument error)
A temperature error occurred. The temperature is always monitored.
Turn back on the instrument and use the instrument within the operating temperature range. Repair is necessary even when the instrument is used in the operating temperature range but an error still occurs.
ON output
99
-383,"Temperature error"
ERR
DDE
Not cleared
-384
Output voltage error (hardware
error)
Output of a voltage higher than the test voltage was detected during testing.
Turn off the instrument and remove the test
leads.
Turn on the instrument and perform a
test. If you encounter an error
, the instrument
needs to be repaired.
ON output
99
-384,
”
Overvolt
error
”
ERR
DDE
Not cleared
-385
Fan error
(instrument error)
A fan error occurred. The fan is always monitored.
Repair is necessary
.
ON output
99
-385,"Fan failed"
ERR
DDE
Not cleared
-387
Corrupted backup data
(instrument error)
Damaged data is detected during a backup data check during the startup.
Repair is necessary
.
ON output
99
-387,"Setting backup lost"
ERR
DDE
Not cleared
-388
Corrupted adjustment data
(instrument error)
Damaged data was detected during an adjustment data check during the startup.
Repair is necessary
.
ON output
99
-388,"Calibration memory lost"
ERR
DDE
Not cleared
-390
FPGA
error
(instrument error)
An error was detected during a communications check between the CPU and FPGA
during the
startup.
Repair is necessary
.
ON output
99
-390,"FPGA failed"
ERR
DDE
Not cleared
-400
Query error
The transmission buffer was
full when a communications command query was sent.
Be sure to receive a response when a query is sent.
-400,"Query error"
ERR
QYE
HIOKI BT5525A961-00
Содержание BT5525
Страница 2: ...HIOKI BT5525A961 00 ...
Страница 6: ...Contents 6 HIOKI BT5525A961 00 ...
Страница 48: ...48 Notifying the Judgment Result and Test Completion with a Beep HIOKI BT5525A961 00 ...
Страница 76: ...76 Initializing the Instrument Reset HIOKI BT5525A961 00 ...
Страница 82: ...82 Deleting Panel Data HIOKI BT5525A961 00 ...
Страница 200: ...200 License Information HIOKI BT5525A961 00 ...
Страница 215: ...www hioki com HIOKI BT5525A961 00 ...