Breathing system problems
.
Symptom
Problem
Solution
Gas scavenging flow is too low
or too high.
Scavenging extract flow problem.
Use a different scavenging extraction
system. Verify flow is within specification.
Filter blockage. Active systems have
a flow indicator.
Replace the filter. Refer to “Remove the
AGSS receiver filter” in the “Advanced
Breathing System Cleaning and
Sterilization” User’s Reference manual.
The bellows fills when the Bag/
Vent switch is set to Bag or the
bag fills when the switch is set
to Vent.
Leak through Bag/Vent switch.
Contact an authorized service
representative to repair the system.
The ventilator does not read the
position of the Bag/Vent switch.
Ventilator or absorber malfunction.
Ventilate manually. Contact an authorized
service representative to repair the
system.
APL valve does not operate
correctly.
APL valve problem.
Replace APL valve seal and diaphragm.
Large breathing system leak.
Bag hose not connected properly.
Make sure the bag hose is connected to
the bag port (below the APL valve).
Absorber canister not installed
correctly.
Reinstall the absorber canister. Make sure
both pins are engaged.
Bellows falls below top of
indicator during the Bellows
assembly test.
Leak in the breathing system.
Check, clean, or reposition the pressure
relief valve. If the problem persists, replace
the pressure relief valve, bellows base, or
bellows assembly.
During PCV-VG mode,
measured tidal volume is less
than set.
Inspiratory pressure is limited to
Pmax minus 5 cmH2O.
Increase the Pmax setting.
7 Alarms and troubleshooting
2067226-001
7-17
Содержание auisys cs2
Страница 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Страница 25: ...Abbreviation Definition VCV Volume control ventilation Vol Volume 1 Introduction 2067226 001 1 13 ...
Страница 32: ...1 20 2067226 001 ...
Страница 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Страница 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Страница 90: ...3 42 2067226 001 ...
Страница 94: ...4 4 2067226 001 ...
Страница 132: ...7 20 2067226 001 ...
Страница 149: ...2 Auxiliary O2 flow control Figure 8 5 Auxiliary O2 flowmeter 8 Setup and connections 2067226 001 8 17 ...
Страница 164: ...9 8 2067226 001 ...
Страница 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Страница 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Страница 262: ...13 20 2067226 001 ...
Страница 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...