AB.74p047
5.
To clean and disinfect the canister, refer to “Absorbent canister
cleaning” in the “Cleaning and Sterilization” manual.
6.
Place a new filter in the bottom of the canister, pour absorbent
into the canister and place a new filter over the absorbent before
closing and locking the cover. Wipe off any absorbent dust.
7.
Align the cover slots with the canister locking tabs and press the
cover down into place. Turn the cover locking ring clockwise to
lock the cover in place. Ensure the cover is properly sealed to
prevent leaks and spillage. Alignment of the arrows helps to
indicate correct assembly.
AB.82p001
The filters must be in place to help prevent dust and
particles from entering the breathing circuit.
--
8.
When replacing the canister, make sure that it is seated properly
on the support pins or in the EZchange canister module before
latching it into place.
WARNING
8 Setup and connections
2067226-001
8-9
Содержание auisys cs2
Страница 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Страница 25: ...Abbreviation Definition VCV Volume control ventilation Vol Volume 1 Introduction 2067226 001 1 13 ...
Страница 32: ...1 20 2067226 001 ...
Страница 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Страница 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Страница 90: ...3 42 2067226 001 ...
Страница 94: ...4 4 2067226 001 ...
Страница 132: ...7 20 2067226 001 ...
Страница 149: ...2 Auxiliary O2 flow control Figure 8 5 Auxiliary O2 flowmeter 8 Setup and connections 2067226 001 8 17 ...
Страница 164: ...9 8 2067226 001 ...
Страница 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Страница 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Страница 262: ...13 20 2067226 001 ...
Страница 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...